| Issue |
When |
Description |
Required Action |
| Air hose has become detached |
Anytime |
Air hose has detached either from the handpiece or the power unit |
Take a picture of the hose and where it has detached from. Fill out this form to be contacted with further instructions. |
| Air hose is damaged |
Anytime |
The Air hose is visibly damaged with a hole, puncture, or crimping, which impedes sufficient airflow. |
The Air hose must be replaced. Take a picture of the damaged area. Fill out this form to be contacted with further instructions. |
| Device not pulsing |
During operation |
The foot pedal is depressed, the fan is turning on, but no pulse is being emitted. |
Fill out this form to be contacted with further instructions. |
| Device not turning on |
Anytime |
The power button is placed in the ON position and nothing is happening. |
- Check if the unit is plugged in.
- Is the unit plugged in?
- If no, plug unit it and turn it on.
- If yes, check if the red, emergency stop button is depressed.
- Is the emergency stop button depressed?
- If yes, release the emergency stop button and turn the unit on.
- If no, replace fuses and try turning the unit on.
- If not turning on after replacing fuses, move the laser to another room and try plugging the unit in a different electrical outlet.
- If the unit still does not turn on, fill out this form to be contacted with further instructions.
|
| Device turned off |
During operation |
The device has suddenly turned off during operation. |
- Turn the unit off.
- Unplug unit. Replace fuses.
- Plug the unit back in and turn it on.
- If the unit turns on and operates normally, continue use.
- If the unit does not turn back on after replacing the fuses, fill out this form to be contacted with further instructions.
|
| Energy appears to be lower than normal |
Anytime |
Feeling, endpoints, or results appear different during routine operation |
Fill out this form to be contacted with further instructions. |
| Fan is not turning on |
During operation |
The foot pedal is depressed and the laser is pulsing, but the cooling fan does not turn on. |
Fill out this form to be contacted with further instructions. |
| Handpiece is hot |
During operation |
Sufficient air cooling is not occurring. |
Fill out this form to be contacted with further instructions. |
| Lens appears damaged |
Anytime |
The inside or outside of any lens is not completely clear but has a visible mark that cannot be removed by the standard cleaning protocol. |
The lens must be replaced. Take a picture of the lens. Fill out this form to be contacted with further instructions. |
| Service light is on |
Anytime |
The device has emitted 1,000,000 pulses and regular service is due soon. |
It is advised to schedule maintenance. Fill out this form to be to schedule service and receive further instructions. |